UV aging test chamber when install should notice following notes.

First: Do not connect the ground wire to the neutral line of the power supply;

Second: the UV aging test box ground wire can not be connected to the trachea or water pipe;

Third: the UV aging test box power supply installation must be operated by a professional electrician; general enterprises have a professional electrician.

Fourth: please confirm whether the capacity of the total switch of the UV aging test box meets the requirements of the equipment. Each device has a current requirement standard and can not be overloaded.

Fifth: Do not share the equipment with other high and low temperature test chamber etc, to prevent the switch overload leading to test failure; and also to avoid the impact of equipment failure in the way;

Sixth: when intall the box shell, shell must be in contact with the ground. And if the test box is not grounded, the leakage will be the risk of electric shock

Seventh: Please connect the ground wire (yellow-green line) of the UV aging test box to the ground terminal so that the leakage switch can cut off the main power supply when the equipment leakage occurs, thus preventing electric shock.